Gulmez, RecepAltunova, MehmetSahin, Ahmet AnilCelik, Omer2024-05-192024-05-1920241708-53811708-539Xhttps://doi.org10.1177/17085381241236560https://hdl.handle.net/20.500.12713/5673Background: This study aimed to evaluate the efficacy and safety of Pro-Glide, a suture-mediated vascular closure device, regarding technical success and complications in patients who had undergone aortic intervention and had previous groin intervention (PGI). Methods: One hundred and thirty-five patients who underwent percutaneous thoracic endovascular aortic repair via the femoral artery and were closed with the Pro-Glide device were analyzed retrospectively. PGI was defined as a history of open surgical access to the femoral artery or wide sheath (>18 F) placement due to endovascular or valvular intervention. The patients were divided into two groups 38 cases with PGI and 97 cases without PGI. Results: The overall success rate of closure of the femoral artery with Pro-Glide was not statistically significant between the two groups (93.8% vs 92.1%, p = .711). Sheath sizes were compared between the groups and PGI (+) group had significantly higher sheath sizes compared to PGI (-) group (24.3 +/- 1.1 F vs 23.8 +/- 1.0 F, p = .011). Three patients in the PGI (+) group and six patients in the PGI (-) group experienced technical failure of the percutaneous femoral approach. Femoral complications were seen after the procedures in four patients in the PGI (+) group and four in the PGI (-) group. The PGI (+) group had a higher complication rate when compared to the PGI (-) group; however, this was not statistically significant (p = .181). Conclusion: The present study was conducted on a significantly larger sample compared to previous studies and the findings suggest that the Pro-Glide vascular closure device is a safe option for patients with a history of PGI and may not be considered as a contraindication.eninfo:eu-repo/semantics/closedAccessVascular Closure DevicesFemoral ScarThoracic Endovascular Aortic RepairPro-GlidePercloseA single center study of the efficacy and safety of Pro-Glide used for closure in thoracic endovascular aortic repair in patients with previous groin interventionArticle38403595WOS:0011763500000012-s2.0-85186607093N/A10.1177/17085381241236560Q2