Elektroretinografi ve görsel uyarılmış potansiyel ölçümlerinde zenon ve led ışık kaynaklarının karşılaştırılması

dc.authoridSacit Karamürsel / 0000-0002-7534-9392en_US
dc.authorscopusidSacit Karamürsel / 7004361830
dc.authorscopusidSemih Taşdelen / 58619719600
dc.authorwosidSacit Karamürsel / O-9371-2019
dc.authorwosidSacit Karamürsel / J-2130-2018
dc.contributor.authorAksu, Serkan
dc.contributor.authorKurt, Adnan
dc.contributor.authorSoyata, Ahmet Zihni
dc.contributor.authorSacar, Kardelen Turku
dc.contributor.authorTaşdelen, Semih
dc.contributor.authorKaramürsel, Sacit
dc.date.accessioned2020-11-04T12:55:37Z
dc.date.available2020-11-04T12:55:37Z
dc.date.issued2019en_US
dc.departmentİstinye Üniversitesi, Tıp Fakültesi, Temel Tıp Bilimleri Bölümüen_US
dc.description.abstractAmaç: Elektroretinografi (ERG) ve görsel uyarılmış potansiyeller (GUP) görsel bir uyarana yanıt olarak retinada veya oksipital korteksteki primer görme alanında oluşan elektriksel etkinliği ölçen ve görme yollarını nesnel olarak değerlendiren önemli tanı araçlarıdır. Elektriksel potansiyellerin oluşturulması için zenon ya da ışık yayan diyot (LED) ışık kaynağı kullanılabilmektedir. Her iki ışık kaynağı da kullanılarak ölçülen ERG ve GUP kayıtlarını aynı örneklemde karşılaştıran bir araştırmaya rastlanmadığı için, bu çalışmada sağlıklı gönüllülerde zenon ve LED ışık kaynağı kullanılarak oluşturulan ERG ve GUP ölçümlerini karşılaştırmak amaçlanmıştır. Gereç ve Yöntemler: Araştırmanın örneklemi 21–30 yaş aralığındaki 31 sağlıklı gönüllüden oluşmuştur. LED ve zenon ışık kaynağı ile her iki gözden ayrı ayrı ERG ve GUP kayıtları alındı. ERG kayıtlarındaki “a,” “b,” ve GUP kayıtlarındaki N2, P2 dalgalarının gecikme süreleri ve genlikleri eşleştirilmiş t-testiyle karşılaştırıldı. Bulgular: LED ışık kaynağıyla elde edilen ERG kayıtlarında zenon ışık kaynağıyla elde edilen kayıtlara göre a (p<0,001) ve b (p<0,001) dalgalarının genlikleri daha düşük bulundu. LED ışık kaynağı ile elde edilen GUP kayıtlarında ise zenon ışık kaynağı ile elde edilen kayıtlara göre P2 dalgasının genliği (p<0,001) daha düşük bulundu. Ancak her iki ışık kaynağı ile elde edilen ERG ve GUP kayıtlarındaki gecikme süreleri arasında fark saptanmadı. Tartışma ve Sonuç: LED ışık kaynağı ile elde edilen ERG ve GUP kayıtlarının, zenon lamba ile elde edilenlere büyük ölçüde benzer bir dalga morfolojisine sahip olduğu görülmüştür. Ancak bazı dalgalarda daha düşük genlikler elde edilmiştir ve bu durumun görme bozukluklarının saptanmasındaki duyarlılık ve özgüllüğünün azalmasına neden olabileceği düşünülmüştür. LED ışık kaynağı ile yapılan GUP ve ERG ölçümlerindeki genlikleri artırmak için monokromatik LED ve özel gözlük kullanılması gibi değişiklikler önerilebilir.en_US
dc.description.abstractAim: Electroretinography (ERG) and visual evoked potentials (VEP) are important diagnostic tools that measure the electrical activity occurring in the retina or the occipital cortex after visual stimuli and assess the visual pathways objectively. Xenon or light emitting diode (LED) light sources can be used to generate the electrical potentials. Because no previous study was found to compare ERG and VEP measurements generated by using both light sources in the same sample, this study aimed to compare ERG and VEP measurements using both LED and xenon light sources in healthy volunteers. Materials and Methods: The study sample comprised 31 healthy volunteers aged between 21 and 30 years. ERG and VEP recordings of both eyes were obtained separately by use of LED and xenon light sources. Latencies and amplitudes of the “a” and “b” waves in ERG and N2 and P2 waves in VEP recordings were compared using the paired t-test. Results: Amplitudes of the “a” (p<0.001) and “b” (p<0.001) waves in the ERG recordings using LED light source were found to be lower than in those using xenon light source. Amplitudes of P2 wave (p<0.001) were found to be lower in the VEP recordings using LED light source than in those using xenon light source. However, no difference was found between the ERG and VEP measurements using each light source in terms of latency. Discussion and Conclusion: Highly similar wave morphologies were observed in the ERG and VEP measurements using xenon and LED light sources. However, lower amplitudes were obtained in some of the waves, which was thought to potentially cause reduced sensitivity and specificity in the detection of visual pathologies. Use of monochromatic LED or special goggles can be suggested for increased amplitudes in the ERG and VEP measurements to use LED light source.en_US
dc.identifier.citationAKSU S,KURT A,SOYATA A. Z,SAÇAR K. T,TAŞDELEN S,KARAMÜRSEL S (2019). Elektroretinografi ve Görsel Uyarılmış Potansiyel Ölçümlerinde Zenon ve LED Işık Kaynaklarının Karşılaştırılması. ANADOLU KLİNİĞİ TIP BİLİMLERİ DERGİSİ, 24(2), 78 - 86. Doi: 10.21673/anadoluklin.472427en_US
dc.identifier.doi10.21673/anadoluklin.472427en_US
dc.identifier.endpage86en_US
dc.identifier.issn2149-5254 / 2458-8849en_US
dc.identifier.issue2en_US
dc.identifier.startpage78en_US
dc.identifier.trdizinid369777en_US
dc.identifier.urihttps://doi.org/10.21673/anadoluklin.472427
dc.identifier.urihttps://hdl.handle.net/20.500.12713/1170
dc.identifier.volume24en_US
dc.indekslendigikaynakTR-Dizinen_US
dc.institutionauthorKaramürsel, Sacit
dc.language.isotren_US
dc.relation.ispartofANADOLU KLİNİĞİ TIP BİLİMLERİ DERGİSİen_US
dc.relation.publicationcategoryMakale - Ulusal Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectElektrofizyolojien_US
dc.subjectElektroretinografien_US
dc.subjectGörsel Testleren_US
dc.subjectGörsel Uyarılmış Potansiyelleren_US
dc.subjectElectrophysiologyen_US
dc.subjectElectroretinographyen_US
dc.subjectVisual Evoked Potentialsen_US
dc.subjectVisual Testsen_US
dc.titleElektroretinografi ve görsel uyarılmış potansiyel ölçümlerinde zenon ve led ışık kaynaklarının karşılaştırılmasıen_US
dc.title.alternativeA comparison of xenon and led light sources in electroretinography and visual evoked potential measurementsen_US
dc.typeArticleen_US

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
26.pdf
Boyut:
670.38 KB
Biçim:
Adobe Portable Document Format
Açıklama:
Tam Metin / Full Text
Lisans paketi
Listeleniyor 1 - 1 / 1
Küçük Resim Yok
İsim:
license.txt
Boyut:
1.44 KB
Biçim:
Item-specific license agreed upon to submission
Açıklama: