Calculation of NaI(Tl) detector efficiency using Ra-226, Th-232, and K-40 radioisotopes: Three-phase Monte Carlo simulation study

Yükleniyor...
Küçük Resim

Tarih

2022

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

DE GRUYTER POLAND SP Z O O

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

Thallium-activated sodium iodide (NaI(Tl)) detectors can be used in gamma cameras, environmental radiation assessments, including radiation emission levels from nuclear reactors, and radiation analysis equipment. This three-phase investigation aimed to model a standard NaI(Tl) detector using the Monte Carlo N-Particle eXtended (MCNPX) general-purpose Monte Carlo simulation techniques. Accordingly, a standard NaI(Tl) detector was designed along with the required properties. Next a validation study of the modelled NaI(Tl) detector has been performed based on the experimental results for absolute detector efficiency values obtained from Ra-226, Th-232, and K-40 radioisotopes. Our findings indicate that the obtained absolute detector efficiency values are quite close to used experimental values. Finally, we used the modelled detector for determination of mass attenuation coefficients of Ordinary concrete, Lead, Hematite-serpentine concrete, and Steel-scrap concrete at 186.1, 295.22, 351.93, 609.31, 1120.29, 1764.49, 238.63, 911.2, 2614, and 1460.83 keV gamma-ray energies. Additionally, according to our findings, mass attenuation coefficients obtained from the newly designed detector are compatible with the standard NIST (XCOM) data. To conclude, continuous optimisation procedures are strongly suggested for sophisticated Monte Carlo simulations in order to maintain a high degree of simulation reliability. As a result, it can be concluded that the validation of the simulation model is necessary using measured data. Finally, it can also be concluded that the validated detector models are effective instruments for obtaining basic gamma-ray shielding parameters such as mass attenuation coefficients.

Açıklama

Anahtar Kelimeler

NaI(Tl) Detector, MCNPX, Absolute Detector Efficiency, Scintillation Detectors

Kaynak

OPEN CHEMISTRY

WoS Q Değeri

Q3

Scopus Q Değeri

Q2

Cilt

20

Sayı

1

Künye

Tekin, H. O., ALMisned, G., Issa, S. A. M., Zakaly, H. M. H., Kilic, G., Ene, A. (2022). Calculation of NaI(Tl) detector efficiency using Ra-226, Th-232, and K-40 radioisotopes: Three-phase Monte Carlo simulation study. Open Chemistry, 20(1), 541-549.