Development of independent mu/treatment time verification algorithm for non-imrt treatment planning: a clinical experience

Yükleniyor...
Küçük Resim

Tarih

2018

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Amer Inst Physics

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

The aim of this study is to develop an algorithm for independent MU/treatment time (TT) verification for non-IMRT treatment plans, as a part of QA program to ensure treatment delivery accuracy. Two radiotherapy delivery units and their treatment planning systems (TPS) were commissioned in Liv Hospital Radiation Medicine Center, Tbilisi, Georgia. Beam data were collected according to vendors' collection guidelines, and AAPM reports recommendations, and processed by Microsoft Excel during in-house algorithm development. The algorithm is designed and optimized for calculating SSD and SAD treatment plans, based on AAPM TG114 dose calculation recommendations, coded and embedded in MS Excel spreadsheet, as a preliminary verification algorithm (VA). Treatment verification plans were created by TPSs based on IAEA TRS 430 recommendations, also calculated by VA, and point measurements were collected by solid water phantom, and compared. Study showed that, in-house VA can be used for non-IMRT plans MU/TT verifications.

Açıklama

33rd International Physics Congress of the Turkish-Physical-Society (TPS) -- SEP 06-10, 2017 -- Bodrum, TURKEY

Anahtar Kelimeler

Kaynak

Turkish Physical Society 33rd International Physics Congress (Tps33)

WoS Q Değeri

N/A

Scopus Q Değeri

N/A

Cilt

1935

Sayı

Künye

Tatli, H., Yucel, D., Yilmaz, S., & Fayda, M. (2018, February). Development of independent MU/treatment time verification algorithm for non-IMRT treatment planning: A clinical experience. In AIP Conference Proceedings (Vol. 1935, No. 1, p. 090001). AIP Publishing LLC.