Positional accuracy of a single implant analog in additively manufactured casts in biobased model resin

dc.authorscopusidMustafa Borga Dönmez / 57202022054
dc.authorwosidMustafa Borga Dönmez / AGY-6155-2022
dc.contributor.authorDiken Türksayar, Almira Ada
dc.contributor.authorDemirel, Münir
dc.contributor.authorPetersmann, Sandra
dc.contributor.authorSpintzyk, Sebastian
dc.contributor.authorDönmez, Mustafa Borga
dc.date.accessioned2025-04-18T10:11:25Z
dc.date.available2025-04-18T10:11:25Z
dc.date.issued2024
dc.departmentİstinye Üniversitesi, Diş Hekimliği Fakültesi, Klinik Bilimler Bölümü
dc.description.abstractObjectives: To evaluate the positional accuracy of implant analogs in biobased model resin by comparing them to that of implant analogs in model resin casts and conventional analogs in dental stone casts. Methods: Polyvinylsiloxane impressions of a partially edentulous mandibular model with a single implant were made and poured in type IV dental stone. The same model was also digitized with an intraoral scanner and additively manufactured implant casts were fabricated in biobased model resin (FotoDent biobased model) and model resin (FotoDent model 2 beige -opaque) ( n = 8). All casts and the model were digitized with a laboratory scanner, and the scan files were imported into a 3 -dimensional analysis software (Geomagic Control X). The linear deviations of 2 standardized points on the scan body used during digitization were automatically calculated on x-, y-, and z-axes. Average deviations were used to define precision, and 1 -way analysis of variance and Tukey HSD tests were used for statistical analyses ( alpha = 0.05). Results: Biobased model resin led to higher deviations than dental stone (all axes, P <= 0.031) and model resin (yaxis, P = 0.015). Biobased model resin resulted in the lowest precision of implant analog position ( P <= 0.049). The difference in the positional accuracy of implant analogs of model resin and stone casts was nonsignificant ( P >= 0.196). Conclusions: Implant analogs in biobased model resin casts mostly had lower positional accuracy, whereas those in model resin and stone casts had similar positional accuracy. Regardless of the material, analogs deviated more towards mesial, while buccal deviations in additively manufactured casts and lingual deviations in stone casts were more prominent.
dc.identifier.citationTürksayar, A. A. D., Demirel, M., Petersmann, S., Spintzyk, S., & Donmez, M. B. (2024). Positional accuracy of a single implant analog in additively manufactured casts in biobased model resin. Journal of dentistry, 146, 105037.
dc.identifier.doi10.1016/j.jdent.2024.105037
dc.identifier.endpage7
dc.identifier.issn0300-5712
dc.identifier.issn1879-176X
dc.identifier.pmid38703808
dc.identifier.scopus2-s2.0-85192303654
dc.identifier.scopusqualityQ1
dc.identifier.startpage1
dc.identifier.urihttp://dx.doi.org/10.1016/j.jdent.2024.105037
dc.identifier.urihttps://hdl.handle.net/20.500.12713/6983
dc.identifier.volume146
dc.identifier.wosWOS:001240061000001
dc.identifier.wosqualityQ1
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.indekslendigikaynakPubMed
dc.institutionauthorDönmez, Mustafa Borga
dc.institutionauthoridMustafa Borga Dönmez / 0000-0002-3094-7487
dc.language.isoen
dc.publisherElsevier
dc.relation.ispartofJournal of dentistry
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectAdditive Manufacturing
dc.subjectBiobased Resin
dc.subjectImplant Analog
dc.subjectImplant Cast
dc.subjectPositional Accuracy
dc.titlePositional accuracy of a single implant analog in additively manufactured casts in biobased model resin
dc.typeArticle

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
1-s2.0-S0300571224002070-main.pdf
Boyut:
7.28 MB
Biçim:
Adobe Portable Document Format
Lisans paketi
Listeleniyor 1 - 1 / 1
Küçük Resim Yok
İsim:
license.txt
Boyut:
1.17 KB
Biçim:
Item-specific license agreed upon to submission
Açıklama: